TF3 12/125/1200 CC-T1 Touchscreen Controller (3 Zones) Medium 3-zone tube furnac

TF3 12/125/1200 CC-T1 Touchscreen Controller (3 Zones) Medium 3-zone tube furnac
La imagen es solo una referencia, puede diferir e incluir accesorios no disponibles en el producto.
  • (Haga clic aquí para obtener más información)
  • pdf_preview
 
Número de artículo.: 00053318 21.815,93 € para 1 Pieza más IVA (21%) más costes de envío
Medium 3-zone tube furnace
Wire heating elements in high quality vacuum formed insulation

1200°C max. temperature
1200 mm heated length
1550 mm tube length for use in air
1800 mm tube length for use with modified atmosphere
125 mm max. outer work tube diameter
430x1465x575 mm (HxWxD) outer dimensions furnace including control box
Digital PID controller (see details in separate position)
220x655x480 mm (HxWxD) outer dimensions control module
1055 mm uniform temperature length +/-5°C at 1100°C
4.46 kW max power rating
Type N thermocouple
134 kg weight

The picture shows a product from the same range. Size and features may differ from this.

The furnace body is supplied separated from the control box, for horizontal use mounted on the bench.
There are 2 m long interconnecting cables between the furnace and control box which can be easily unplugged at the back of the control box.

temperature programmer for three zone products (with retransmission of setpoint control system)

CC-T1 Programmer and data-logger with USB and ethernet connection

- Three heating zone control
- 4.3” color touch screen
- 10 unique program profiles saved in memory
- 24 segments per unique program (e.g. 12 ramps and 12 dwell times)
- Max. 2 relay operated options (e.g. solenoid valves, if fitted optionally)
- USB communication port at the front of the control box
- Ethernet port at the back of the control box
- Data logging to csv file on customer’s USB stick
- Real time clock and calendar
- Program schedule start
- Program status indication with estimate end time & date
- Control power indication
- Language setting: English, German, French, Italian, Spanish, Chinese
- User level security with three levels:
1. Operator can change temperature setpoint / run pre-configured programs / run data logging
2. Supervisor can additionally configure and edit programs / manage alarms
3. Administrator can additionally set language, time and date / edit data logging settings / edit settings including calibration and serial communication

To run the furnace, please order one of the following packages
- Air package IAP 75x86x1550mm
- Inert gas package IAP 75x86x1800mm
- Vacuum package (on request)

Optional
Digital over-temperature protection (00053325)
+ info